CPU/NIBP Module Assembly with LCD Display
4. Transmits/Receives data to/from Infrared/Predictive Temperature modules
5. Manages Recorder functions
6. Manages Keypad interface
7. Provides RS232 communication to PC for software download
8. Supports Service Diagnostics
9. Supports Nurse Call feature
10. Checks error conditions
11. Transmits/Receives data to/from SpO
12. Controls LCD view angle
1.2.2
CPU Hardware Controls
1. Microcontroller executes the above software functions, including:
2. Battery Back Up Circuitry for Power Savings
3. Flash ROM for Program Storage and Execution
4. Static RAM for Data Storage
5. RS232 level translation for PC Download and second interface for CIS/HIS
6. Real time clock with 16K of Lithium backed up NV RAM
1.2.3
NIBP Hardware functions
1.2.3.1
Main pressure transducer circuitry
A pressure transducer and amplifier combination is provided to acquire the cuff pressure
signal.
1.2.3.2
Over pressure transducer circuitry
An over pressure transducer, independent of the main pressure transducer. Over pressure
limits are set according to the patient size.
1 - 4
•
Maintain watchdog timer
•
Run-time diagnostic
•
NIBP
•
Recorder
•
Temperature module, PTM or IR
•
SpO
2
•
LCD
•
Static memory
•
Non-volatile RAM for configuration and future serial number
0070-10-0691
module, if present
2
Theory of Operation
Accutorr Plus Service Manual
Need help?