7.9.6

Test display

How to get there
Press the
press the
What will be tested ?
The test includes all segments of the LC display, the background illumina-
tion and all light emitting diodes of the DLX. First the illumination of the dis-
play will be switched off and all LEDs will be lit. Then the display
illumination is switched back on and all segments are switched to black. Fi-
nally all light emitting diodes are lit for a short time, one after another.
The display test will take approximately 10 seconds. Once it is complete,
the DLX returns to the menu item "Maintenance"
The display test has no effect on the measurement or the state of the out-
puts.
7.10

Main menu item "Erase Alarms"

Figure 41, Flow diagram for menu item "Erase alarms"

How to get there
Press the
Erasing alarms
The user can erase all alarms that have occurred. Possibly active alarm in-
dication relays will be reset and the light emitting diodes for alarm indication
will stop blinking.