Service Interface
The service interface underneath the keypad can be used to read or modify
register contents via the programming software DLXPARA (after input of a
password). This is a serial connection (8, e, 1) with fixed baud rate (9600
baud) using a standard serial cable (e.g. modem cable).
All register addresses have 5 digits plus a 2 digit index. The following table
lists all of the registers. Use the keypad to query these values directly.
Meaning of special symbols in the table
The column „Stat." Shows whether this register can be set or programmed
in a certified and sealed unit:
S
Value can be
set
word)
P
Value can be
programmed
tection switch is set to „Set Enable"), these values cannot be modified in a
certified and sealed unit!
All locations marked with gray background indicate registers that cannot
be modified from outside in a certified and sealed unit!
(i.e. it can be modified at any time after entering a pass-
(i.e. it can only be modified if the program pro-