1.2.14 The radiometer performs measurements under the following conditions:
- temperature from – 20 to + 50 o C;
- relative humidity up to (95±3) % at + 35 o C temperature;
- atmospheric pressure from 84 to 106.7 kPa.
1.2.15 The radiometer is resistant to the following external factors:
- high frequency sinusoidal vibrations (with crossover frequency from 57 to 62 Hz)
in the range of 10 to 55 Hz, 0.15 mm bias for frequency lower than the crossover
frequency;
- shocks with a shock pulse duration of 5 ms, a total number of shocks 1000±10
and maximum acceleration of 100 m/s
- shocks in shipping container with an acceleration of 98 m/s
pulse duration of 16 ms (number of shocks - 1000±10 in each direction), or
equivalent shake tests;
- exposure in shipping container to ambient air temperature from - 25 to
+ 55 o C and relative humidity up to (95±3) % at + 35 o C;
- photon-ionizing radiation with exposure dose rate that corresponds to ambient
DER up to 0.1 Sv/h during 5 minutes.