Electrical Connection Sensor Connecting Cable; Prosonic Flow 91; Supply Voltage (Power Supply); Cable Entry - Endress+Hauser Proline Prosonic Flow 90U Technical Information

Ultrasonic flow measuring system; flowrate measurement for standard applications with drinking water, wastewater and process water
Table of Contents

Advertisement

Electrical connection
sensor connecting cable

Prosonic Flow 91

Supply voltage (power supply)

Cable entry

22
Connecting power sensor cables in the connection compartment
a
b
c
Connecting the measuring system
a, b
Sensor connecting cables
c
Cover of cable gland
d
Rubber seal
e
Cable retaining sleeves
f
Ground disk
g
Cable gland holder
h
Seal
i
Cable holder
Transmitter:
• 85 to 260 V AC, 45 to 65 Hz
• 20 to 55 V AC, 45 to 65 Hz
• 16 to 62 V DC
Measuring sensors:
• Powered by the transmitter
Power supply and signal cables (inputs/outputs):
• Cable entry M20 x 1.5 (8 to 12 mm)
or
• Cable gland for cables with ∅ 6 to 12 mm
• Threaded adapter ½" NPT, G ½"
Sensor connecting cable:
A special cable gland allows you to insert both sensor cables (per channel) into the connection compartment
simultaneously.
• Cable gland M20 x 1.5 for 2 x ∅ 4 mm
or
• Threaded adapter ½" NPT, G ½"
Proline Prosonic Flow 90U, 90W, 91W, 93C, 93U, 93W