Repeatability - Endress+Hauser Proline Prosonic Flow 90U Technical Information

Ultrasonic flow measuring system; flowrate measurement for standard applications with drinking water, wastewater and process water
Table of Contents

Advertisement

Proline Prosonic Flow 90U, 90W, 91W, 93C, 93U, 93W

Repeatability

Endress + Hauser
Max. measured error for dry calibration in % of reading
%
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.0
0
2
a
Pipe diameter > DN 200
b
Pipe diameter > DN 50 < DN 200
c
Pipe diameter < DN 50
Max. measured error for wet calibration and verification of accuracy in % of reading
%
4.0
3.5
3.0
2.5
2.0
1.5
1.0
0.5
0.0
0
2
a
Pipe diameter > DN 200
b
Pipe diameter > DN 50 < DN 200
c
Pipe diameter < DN 50
± 0.3% for flow velocities > 0.3 m/s
a
b
c
4
6
8
m/s
a
b
c
4
6
8
m/s
10
12
14
10
12
14
A0003718-en
A0003720-en
25