Table 17. Guidance And Manufacturer's Declaration - Electromagnetic Immunity - Dentsply Sirona Schick 33 User Manual

Usb module and sensors
Table of Contents

Advertisement

Table 17. Guidance and Manufacturer's Declaration - Electromagnetic Immunity

PLEASE NOTE: The USB Interface is intended for use in the electromagnetic environment specified below.
The customer or user of the USB Interface must ensure that it is used in such an environment.
I
T
MMUNITY
EST
Electrostatic discharge
6 kV contact
(ESD)
IEC 61000-4-2
8 kV air
Electrical fast
2 kV for power
transient/burst
supply lines
IEC 610004-4
1 kV for
input/output lines
Surge
 1 kV differential
IEC 61000-4-5
mode
 2kV common
mode
Voltage dips, short
< 5% U
interruptions and
(>95% dip in
voltage variations on
for 0.5
power supply input
lines
40% U
IEC 61000-4-11
(60% dip in
for 5
70% U
(30% dip in
for 25
< 5% U
(>95% dip in
for 5
Power frequency
3A/m
(50/60 Hz) magnetic
field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
74
IEC 60601 T
C
EST
L
EVEL
6 kV contact
8 kV air
2 kV for power
supply lines
1 kV for
input/output lines
 1 kV differential
mode
 2kV common
mode
< 5% U
T
U
T)
(>95% dip in
cycle
for 0.5
40% U
T
U
T)
(60% dip in
cycles
cycles
for 5
70% U
T
U
T)
(30% dip in
cycles
for 25
< 5% U
T
U
T)
(>95% dip in
sec
sec
for 5
3A/m
3 Vrms
100003870 Rev. 4
OMPLIANCE
L
EVEL
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
a typical commercial or hospital
environment.
Mains power quality should be that of
T
U
a typical commercial or hospital
T)
cycle
environment. If the user of the USB
Interface requires continued operation
during mains interruptions, it is
T
U
recommended that the PC workstation
T)
to which the USB Interface is
connected be powered from an
uninterruptible power supply or
T
U
battery.
T)
cycles
NOTE: U
prior to application of the test level.
T
U
T)
Mains power quality should be that of
a typical commercial or hospital
environment.
Portable and mobile RF
communication equipment should be
used no closer to any part of the USB
Interface, including its cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d 
1.2
P
Schick USB Module and Sensors
G
UIDANCE
is the AC mains voltage
T

Hide quick links: