7.4.
USB Interface Firmware Upgrades
Schick USB Module and Sensors
Field updates to the USB Interface can be accomplished by installing
new firmware.
Perform the following steps to upgrade USB Interface firmware.
S
TEP
1
IMPORTANT! Close CDR DICOM (if running).
2
Verify that USB Interface is connected to the host computer.
Start the Upgrade Utility (Start > All Programs > CDR DICOM
3
for Windows > Upgrade Utility).
If the firmware version number is listed in red, click the Firmware
Version button to upgrade. The firmware version number will
4
change to green letters when the upgrade is completed
successfully.
If the FPGA version number is listed in red, click the FPGA
Version button to upgrade. The firmware version number will
5
change to green letters when the upgrade is completed
successfully.
Verify that the USB Interface firmware and FPGA version
6
numbers are listed in green. If not, please contact your distributor
of Dentsply Sirona products for additional information.
7
Click Close to exit the Upgrade Utility.
100003870 Rev. 4
When
new
firmware
available,
it
provided with software release
and becomes part of the
update
to
your
system.
In the event that version
information is displayed with
red lettering, we recommend
that you perform the upgrade,
referring to the procedure
below.
A
CTION
is
is
typically
existing
53