3. LED Indicators
3.1.
Schick 33/Elite USB Interface
Table 2. Schick 33/Elite USB Interface LED Indications (Normal Conditions)
CDR
S
ENSOR
DICOM
Connected
Running
Connected
Running
Not
Connected
running
Not
Not
connected
running
Table 3. Schick 33/Elite USB Interface LED Indications (Error Conditions)
CDR
S
ENSOR
DICOM
Not
Connected
running
Not
Connected
running
10
S
R
ENSOR
EADY
S
A
TATUS
CQUIRE
Flashing every
ON
½ - 1½
seconds
ON
OFF
OFF
OFF
OFF
ON
S
R
ENSOR
EADY
S
A
TATUS
CQUIRE
OFF
ON
OFF
ON
100003870 Rev. 4
S
ENSOR
T
O
C
ONNECTED
ON
ON
ON
OFF
S
ENSOR
T
O
C
ONNECTED
OFF
Flashing
Green
Schick USB Module and Sensors
D
ESCRIPTION
Ready for image
acquisition
Image acquisition
Start CDR DICOM
to prepare for image
acquisition
Connect Sensor and
start CDR DICOM
D
ESCRIPTION
Cable short or over-
current condition.
Replace Sensor
cable. Replace
Sensor if problem
persists.
Low current
condition. Replace
Sensor cable.
Replace Sensor if
problem persists.