Table 6. Sensor Positioning (AimRight Grip System)

E
P
I
XAMPLES
LACEMENT
NSTRUCTIONS
Maxillary Anterior
Place the distal end of the sensor against the
roof of the mouth, with the incisal edge of the
teeth against the front of the tab.
Sensor should be parallel to the long axis of
the maxillary anterior teeth. Ensure the ring is
as close to the patient's face as possible and
place the x-ray head against the ring.
Mandibular Anterior
Place the Sensor into the lower anterior area,
positioning it on top of the tongue, parallel to
the first molar.
Sensor should be centered on the mandibular
anterior teeth when the patient is occluded.
Ensure the ring is as close to the patient's face
as possible and place the x-ray head against the
ring.
Horizontal Bitewing
Place the Sensor between the tongue and the
teeth with the bite area resting on the premolar
teeth.
The patient should close on their back teeth to
ensure centric occlusion and as they do so, the
arm should be angled gently toward the
midline of the mouth to ensure the Sensor is
parallel with the teeth and to provide open
contacts. Ensure the ring is as close to the
patient's face as possible and place the x-ray
head against the ring.
17
Schick USB Module and Sensors
100003870 Rev. 4