7.5.

Sensor Module Firmware Upgrade

54
Field updates to the Sensor can be accomplished by installing new
firmware.
Perform the following steps to upgrade Sensor firmware.
S
TEP
1
IMPORTANT! Close CDR DICOM (if running).
Verify that the Sensor is connected to the USB Interface host
2
computer.
Start the Upgrade Utility. (Start > All Programs > CDR DICOM
3
for Windows > Upgrade Utility).
If the firmware version number is listed in red, click the Firmware
Version button to upgrade. The firmware version number will
4
change to green letters when the upgrade is completed
successfully.
Verify that the Sensor Module firmware version number is listed
5
in green. If not, please contact your distributor of Dentsply
Sirona products for additional information.
6
Click Close to exit the Upgrade Utility.
.
100003870 Rev. 4
When
new
firmware
available,
it
is
provided with software release
and becomes part of the
update
to
your
system.
In the event that version
information is displayed with
red lettering, we recommend
that you perform the upgrade,
referring to the procedure
below.
A
CTION
Schick USB Module and Sensors
is
typically
existing