ITEM
Technology
Pixel Size
Line Pairs
Active Sensor
Area
External
Dimensions
Cable Length
Transport and Storage
Conditions
Operating Conditions
Operating Altitude
Restricted service
statement

A-4. Leakage Current Statement

72
CMOS-APS (Active Pixel Sensor)
15 µm (native); 30 µm (effective) pixel resolution
16.7 lp at 30 µm
Size 0
18 x 24 mm
Size 1
20 x 30 mm
Size 2
26 x 36 mm
Size 0
23.5 x 32 x 6.3 mm
Size 1
25.3 x 38.4 x 6.7 mm
Size2
31.2 x 43.9 x 6.3 mm
2.70 m (maximum)
Ambient temperature range: -40 F (-40 C) to 158 F (+70 C)
Relative humidity range: 10 to 100%, including condensation
Atmospheric pressure range: 500 hPa to 1060 hPa
Ambient temperature range: 50 F (+10 C) to 104 F (+40 C)
Relative humidity range: less than 75%
Atmospheric pressure range: 700 hPa to 1060 hPa
< 2000 m
Unless otherwise specified, this unit should be serviced only by the manufacturer.
It contains no user-serviceable parts.
The system complies with the leakage current requirements of IEC
60601-1 safety standard. Variations, however, may exist in the
construction of computers to which the USB Interface is connected.
Customers are advised to have a qualified electrician perform a leakage
test on their equipment before using the USB Interface.