9.3.

Recommended Disinfectant

Schick USB Module and Sensors
6. If the Sensor or cable are visibly soiled (e.g., with blood or
saliva), each should be cleaned with a soapy cloth or paper
towel, and then dried with a clean lint-free cloth or paper
towel.
7. Thoroughly spray or wipe the Sensor and cable with one of the
disinfecting products recommended in Section 9.3 on page 61.
Do not expose the Sensor cable connectors to any amount of
liquid.
8. If using a spray disinfectant, allow it to remain on the Sensor
for 5 minutes. If using a liquid disinfectant, allow it to remain
on the Sensor for 30 seconds.
9. Repeat steps 7 and 8. When the Sensor has been sprayed or
wiped two times, continue with the following steps.
10. Remove potential chemical build-up from the Sensor by wiping
it with a lap sponge saturated with de-ionized water.
11. Use a dry lap sponge to dry the Sensor or cable, as needed.
12. Place the Sensor in a clean environment, ready for next use.
13. Reconnect the Sensor.
14. Remove and discard gloves.
The following surface disinfectant has been found to be effective in
achieving a desired level of disinfection and is available from other
suppliers.
 Cavi-Wipes (Metrex Research, Kerr) or equivalent
100003870 Rev. 4
61