Special measures

After collision

• Eliminate the cause of the collision.
• Switch on the drives.
• Turn the rotary switch on of the control cabinet clockwise and
hold it until the indicator lamp located above lights up.
• Initialize the CMM if the cause cannot be eliminated by means of
the above measure.
Operating instructions for the measuring software.
• Then switch on the drives.
Collision during probe change
A collision occurring during probe change may be caused by missing
intermediate positions.
• Reallocate the storage position to the probe and choose the inter-
mediate positions so that probe change can be carried out without
collision.

Probe head outside the measuring volume

If the probe head moves in the z axis outside of the measuring vol-
ume, the drives are switched off. The drives can only be switched on
again after the probe head has been positioned inside the measuring
volume.
Returning the probe head to the measuring volume
To return the probe head to the measuring volume, proceed as fol-
lows: