Calibration Procedure - Zeiss CenterMax Operating Instructions Manual

Coordinate measuring machine for series production
Table of Contents

Advertisement

NOTE
hss
• Screw the metal plate on the measuring table (max. torque
40Nm).
Select the metal plate position so that probing of the probe sphere
is possible with all probes.
• Place the calibration standard on the metal plate and turn the
rotary lever 90° to position I in order to activate the magnet.
The calibration standard can be placed in any position on the metal
plate.
Carrying out a reference measurement
The position of the calibration standard must be calibrated prior to
the length of the styli. Use the master probe for this purpose. Proceed
as follows:
1 Insert the master probe in the adapter plate receptacle
page 6-16.
2 Call the menu for reference measurement
tions for the measuring software.
3 Select probe (stylus) 1 on the control panel.
• Press the button between the five LEDs until the LED for probe
1 lights up.
2
4
5
1
4 Probe the sphere of the calibration standard to determine the
diameter of the sphere.
5 Remove the master probe from the adapter plate receptacle
➤"Removing the probe (manually)" on page 6-17.
Now, you can calibrate all probes.

Calibration procedure

After the reference measurement has been carried out, the probes
can be calibrated.
Semi-automatic calibration should be carried out for all probe sys-
tems.
Tensor calibration should be carried out for the VAST measuring
probe system. This is important since the measuring force causes a
61211-1020202
CenterMax Operating Instructions
Preparations for a measuring run
3
Operating instruc-
7-19