Start-up
Clamping the workpiece on the measuring
table
!
!
!
6-2

Safety measures

Danger!
You are exposed to the danger area of the CMM when working on
the measuring table. Certain safety measures have to be observed to
avoid injuries.
• Switch off the drives to prevent unintentional travel movements.
• Activate the "crane interlock" when using lifting equipment to
place the workpiece on the measuring table.
For more information see
ing software.
Danger!
There must not be any grease film on the measuring table. Work-
pieces may slide and fall off of the measuring table, thus endangering
personnel. Make sure that the measuring table is free of lubricants
before placing a workpiece on top of it.
• Remove any greasy or oily substances from the measuring table.
Important!
The workpieces must have full contact with the measuring table. Any
hard particles lying underneath the workpiece may lead to inaccurate
measurements. Furthermore, the surface of the measuring table and
the workpiece may be damaged.
• Clean the measuring table before placing a workpiece on the mea-
suring table.