PLANAR R54x2
Object Type
Target
Description
Syntax
Related Commands
Equivalent Softkeys

SCPI.TRIGger.SEQuence.SINGle

Method
Instrument
Generates a trigger, independently of the trigger source setting (except for
the External). If the trigger source is set to External, an error occurs (error
code 221) and the command is ignored.
If the instrument is not in the waiting for a trigger state (sweep is in progress
or all the channels are set to Hold), an error occurs (error code 211) and the
command is ignored.
The method does not return control before the end of the sweep (waiting for
the completion of the sweep of all the channels).
app.SCPI.TRIGger.SEQuence.SINGle
SCPI.TRIGger.SEQuence.SOURce
SCPI.INITiate(Ch).CONTinuous
SCPI.INITiate(Ch).IMMediate
None