Scpi.sense(Ch).Correction.collect.ckit.stan(Std).C1 - Planar R54 Programming Manual

Vector reflectometer; programming com/dcom
Table of Contents

Advertisement

PLANAR R54x2

SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1

Object Type
Data Type
Target
Description
Range
Out of Range
Preset Value
Unit
Syntax
Equivalent Softkeys
Property (read/write)
Double
Standard Std of the calibration kit specified for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
Std:
standard number (see Table 7 on page 167)
The C1 value of the open calibration standard.
from –1E18 to 1E18
Sets the value of the limit, which is closer to the specified value.
Varies depending on the selected calibration kit and the standard.
1E–27 F/Hz (Farad/Hertz)
Dim Value As Double
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C1 = 100
Calibration > Edit Cal Kit > Capacitance > C1 [10
COM/DCOM Programming manual
–27
F/Hz]
174