PLANAR R54x2

SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0

Property (read/write)
Object Type
Double
Data Type
Standard Std of the calibration kit specified for channel Ch,
Target
The offset Z0 value of the calibration standard.
Description
from –1E18 to 1E18
Range
Sets the value of the limit, which is closer to the specified value.
Out of Range
50 or 75, depending on the selected calibration kit.
Preset Value
Ω (Ohm)
Unit
Dim Value As Double
Syntax
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 = 50
Calibration > Calibration Kit > Edit Cal Kit > Offset Z0
Equivalent Softkeys
Ch:
channel number 1–4 (see Table 1 on page 22)
Std:
standard number (see Table 7 on page 167)
COM/DCOM Programming manual
185