Scpi.sense(Ch).Correction.collect.acquire.thru - Planar R54 Programming Manual

Vector reflectometer; programming com/dcom
Table of Contents

Advertisement

PLANAR R54x2

SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU

Object Type
Data Type
Target
Description
Range
Out of Range
Syntax
Notes
Equivalent Softkeys
Property (write only)
Variant (Long array)
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
Measures the calibration data of the thru standard between the source port
and the receiver port.
The array contains 2 elements:
Data(0)
the number of the receiver port;
Data(1)
the number of the source port.
Port number is 1 or 2. The array elements can not contain the same port
numbers.
If an incorrect port number is specified, an error occurs (error code: 222). If
the same port numbers are specified, an error occurs (error code: 220).
app.SCPI.SENSe(Ch).CORRection.COLLect.ACQuire.THRU= Array(1, 2)
The property writing starts the measurement for the channel independently
of the trigger initiation and trigger source settings. The function of the
property writing waits for the completion of the measurement.
Calibration > Select Ports > 2-1 (S21 S11) | 1-2 (S12 S22)
Calibration > Thru