Scpi.sense(Ch).Correction.collect.method.response.open - Planar R54 Programming Manual

Vector reflectometer; programming com/dcom
Table of Contents

Advertisement

PLANAR R54x2

SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN

Object Type
Data Type
Target
Description
Range
Out of Range
Syntax
Equivalent Softkeys
Property (write only)
Long
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
Selects the port and sets the response calibration (Open) type for the
calculation of the calibration coefficients on completion of the calibration
executed by the
SCPI.SENSe(Ch).CORRection.COLLect.SAVE
from 1 to 2
An error occurs. Error code: 222.
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.RESPonse.OPEN =
1
None
COM/DCOM Programming manual
method.
187

Advertisement

Table of Contents
loading

Table of Contents