Scpi.initiate(Ch).Immediate - Planar R54 Programming Manual

Vector reflectometer; programming com/dcom
Table of Contents

Advertisement

PLANAR R54x2
Object Type
Target
Description
Syntax
Notes
Equivalent Softkeys

SCPI.INITiate(Ch).IMMediate

Method
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
Sets the channel to the single trigger mode. Before this method is called, the
channel must be in hold state, otherwise an error occurs (error code 213) and
the method is ignored.
On completion of the sweep, the channel goes back into the hold state.
The method returns control before the end of the sweep.
app.SCPI.INITiate(Ch).IMMediate
The sweep start in the single trigger mode depends on the trigger source. If
the trigger is set to internal, the sweep will start immediately after the
method is called. If the trigger is set otherwise, the sweep will start when the
trigger signal is received.
System > Trigger > Single
COM/DCOM Programming manual
142

Advertisement

Table of Contents
loading

Table of Contents