PLANAR R54x2
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Object Type
Target
Description
Syntax
Equivalent Softkeys
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
Object Type
Data Type
Target
Description
Range
Out of Range
Preset Value
Syntax
Equivalent Softkeys
Method
Calibration kit, selected for channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
Resets the calibration kit to the factory settings.
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.RESet
Calibration > Calibration Kit > Edit Cal Kit > Restore
Property (read/write)
Long
Channel Ch,
Ch:
channel number 1–4 (see Table 1 on page 22)
The selected calibration kit for the channel.
from 1 to 14
An error occurs. Error code: 222.
1
Dim Value As Long
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.SELect = 3
Calibration > Calibration Kit
COM/DCOM Programming manual
171