Technical Data; Unit Data; Transport, Storage, And Operating Conditions - Dentsply Sirona Orthophos E Installation Requirements

Hide thumbs Also See for Orthophos E:
Table of Contents

Advertisement

Dentsply Sirona
Installation Requirements Orthophos E
+40
95
106
+10
10
70
kPa
+70
95
106
°C
-40
10
50
kPa
67 10 532 D3352
D3352.021.04.01.02 03.2019
5.2

Technical data

5.2.1

Unit data

Model name:
Nominal voltage:
Permissible fluctuation:
Permissible drop under load: 10%
Rated current:
Nominal frequency:
Mains resistance:
Main building fuse/circuit
breaker:
Power consumption:
5.2.2

Transport, storage, and operating conditions

Orthophos E
Transport and storage
conditions:
Operating conditions:
Operating altitude:
Cephalometer
Transport and storage
conditions:
5 Dimensions, technical data
5.2 Technical data
Orthophos E / Ceph
200 – 240 V
± 10%
max. 12 A
50 Hz / 60 Hz
max. 0.8 ohms
LS B25A slow-blow;
for single connection: B16A/B20A slow-
blow
max. 2.0 kW
Temperature: -10 °C – +70 °C
(14 °F – 158 °F)
Relative humidity: 10% – 95%
Ambient temperature: +10°C – +40°C
(50°F – 104°F)
Relative humidity: 10% - 95% (no
condensation)
≤ 3,000 m above sea level
Temperature: -40°C – +70°C (-40°F –
158°F)
Relative humidity: 10% – 95%
Air pressure: 500 hPa – 1060 hPa
39

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Orthophos e ceph

Table of Contents