7.3.9 Update System time at device

The system time (device-internal permanent calendar with date and time) is saved on a battery-
buffered clock module. Therefore the time information or the system time is also available after
switching OFF / ON the device. Update the system time is the upper term for the synchronization
and the setup of date and time. The system time can be set in the "Date & Time" display. For this
purpose you need the corresponding rights and the security switch (SSW) must be switched
according to the parameterization. Synchronization enables the Time Synchronization display
and can be carried out by all users.
Figure 7.60: Displays to update the system time on the device or on remote operation panel
Time synchronization
Procedural steps for
1.
Open the Time Synchronization display (System >Time Service > Date & Time) without login.
Enter the seconds (field after Time synchronization by) to synchronize the system time.
Positive and negative values are allowed. The field only accepts values within the
2.
parameterized synchronization window.
Check the status of "Next possible synchronization". If Now is shown you can
3.
synchronize the system time (step 4). Otherwise, wait until the specified time is reached.
Activate the Synchronize now action, the expected new time is taken as the system time
4.
For further information on the settings and parameter settings of this function, see chapter 8 and
enSuite Online Help.