Technical data and notes

10
Technical data and notes
The essential technical data is shown on the various type plates and labels directly on the device and are
reproduced here below. Supplementary tables and the CE Declaration complete this overview.

10.1 Tables and information additional to the device

Q2 System
Location of use
Mounting clearance
Gas group
Gas ports */
Gas connection
Breather vent line
Gas management
Size of housing /dimensions
Weight
IP rating
Voltage / Curent / Power
Operation panel
Heat-up time
Housing heater
Dynamics
Operational calibration interval
settin
Service calibration interval
Data memory (non-volatile):
Operating system
Software
File Format:
Data logger
*Standard equipment other connections available on request
GasLab Q2
Values and / or Comment
Installation close to the sampling point up to Zone 1
left and right at least 300 mm / top and bottom at least 200 mm / free space
(recommended 900 x 900 mm)
only gases which are according to appropriate gases below
IIC / (Group: A B C D pending)
It is not allowed to supply gasses which can react in the abcence of air!
Input (1 x Process gas. 1 x calibration gas mixture)
Output (1 x Process gas bypass 1 x Measurement gas)
Swagelok fittings 6 mm or ¼ " could be supplied by Elster
Breather connection: 1/4 "NPT
Integrated 2-channel double block and bleed device
for process gas and calibration gas.
H = 38 cm, W = 27 cm, D = 23 cm
Approximately 16kg
IP64
24 V DC Fluctuation, including mains fluctuation ±15% / 4 A / 96 W
Color display 9,5 x 10,5 cm (illuminated); 7 key captive touchpad
2 multi-colored LEDs for indicating the power supply and status
At -25 ° C up to 1 h
30 W
Analysis rate f=1 Hz, reaction time t90≤6s
(90% step response / Tested @ NMi)
1 Week
1 Year
Archives / Calibration data / Parameterization / Logbooks
Service PC: Refer to the download area of Elster
enSuite
enSuite.csv / enSuite.par
Integrated (Logger interval adjustable)
G a s Q u a l i t y A n a l y z e r