GasLab Q2 displays and operating the device
In addition, the logged in user can change his password. The display also provides information
whether a Remote operation panel is active. Only one user can be logged on to a device at a
time. Either the Users Main display shows the Login action or the User (name SU3 in the right
example above) that is already logged in on the device is displayed. This case occurs e.g. during
the transfer of a parameterization to the device.
More information  See chapter 7.3.5
The Logbook (Error List Logbook see figure below) shows stored
Whenever an event becomes active or ends, its state change is recorded in the
logbook. The logbook can contain up to 10000 records; if the logbook is full, the
next entry overwrites the oldest record.
Each event is represented individually and numbered on the right side. The small arrow in front
of the date indicates the beginning of the event (top to bottom) or end of the event (top up). See
following table.
Identification
< black font color>
<yellow font color>
<red font color>
GasLab Q2
Figure 7.25: Display Error List Logbook
Definition
Signal / Message is coming
Signal / Message is going
Signal / Message type event
Signal / Message type warning
Signal / Message type alarm
G a s Q u a l i t y A n a l y z e r
Incidents
Filter selection (same as in the
"Error List Main Display")
Area for displaying errors
(scrollable if more than 2 entries are
contained in the logbook)
Sequence number of logbook
entry