Sony Multiscan CPD-1304S Service Manual page 105

Hide thumbs Also See for Multiscan CPD-1304S:
Table of Contents

Advertisement

Operation and parameter setting on the PC with enSuite
The possibility of switching the parameter type depends on the respective
parameter. For more information about types, units and windows of
parameters, see online help.
During editing, the program uses colored markers to indicate the status. If the parameter label is
blue, this means that the parameter was edited but the change has not yet been saved. Any
superordinate branches are also blue at the same time. A red parameter label indicates that
changes have been made and the parameterization is not yet valid. The parameterization can be
saved, but it can't be transferred to a device.
If you delete a parameter setting an input value or change it fundamentally,
which is used in a other context, all the links are not valid for this value and
are automatically deleted from the parameterization program. You will receive
a message about this deletion so that it is obvious which links must be
recreated.
If you have finished making changes, save the parameterization
disk icon will be grayed out and displayed the lines are in black color. Now you can transfer the
valid and stored device parameterization to the device, with the matching serial number.
for transferring parameterization to device
Working steps
Establish a data connection to the appropriate device. (  See chapter 8.1.2)
1.
Highlight the desired device parameterization in the data structure
2.
of the navigation window
Select the action Transfer parameterization to device (e.g. at the context menu)
3.
Subscribe to the appearing login dialog
4.
The parameterization is transferred. If the new device parameterization is likely to bring about
changes to such parameters, which are protected by the security switch (SSW), but the security
switch is closed, then the message "Device could not be parameterized" appears following the
transfer. The device does not reboot however; the old parameterization is still used.
If the parameters are not under the protection of the security switch or it is open, the transfer is
done and the device reboots. With the reboot, the connections to the remote operation panel and
to enSuite are disconnected and an automatic logout is performed. The device now operates with
the transferred settings. These remain in use even after switched off and on the device.
GasLab Q2
Figure 8.13: Example parameters during editing
G a s Q u a l i t y A n a l y z e r
. After successful storage the
Page 105 of 135
8

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents