Display Control (Calibration / Verification / Touch) - Sony Multiscan CPD-1304S Service Manual

Hide thumbs Also See for Multiscan CPD-1304S:
Table of Contents

Advertisement

GasLab Q2 displays and operating the device

7.2.5 Display Control (calibration / verification / touch)

Behind this icon is the following page which enables you to select "Calibration",
"Verification" or "Disable touch".
Figure 7.17: Display Q2 Control with Sub-displays Q2 Calibration and Q2 Verification
Displays and function of "Operational calibration"  Chapter 7.3.3.
Branching possibilities and function of "Verification"  See chapter 7.3.4
The displays "Service calibration" and "Base calibration" are only required
by Elster and not described any further here.
The display Calibration data contains internal
values caused by the operational calibration
and deviations to the previous settings.
Figure 7.18: Display Q2 Calibration data and Display Q2 Calibration gas
In normal operation, the display Calibration data is not required as it is only used for servicing and
maintenance purposes and facilitate the error analysis. The information on the gas components
is only required when the gases are changed and explained in more detail in chapter 7.3.3.
GasLab Q2
Calibration enables a further branch to the
calibration sub display.
Verification allows the start of the Verification
and other branches.
Disable touch interrupts the operator input
(local and remote) for the count down time
shown behind the link. It can be used, for
example, to prevent temporary inputs caused
by screwing on the sand protection.
The displays Gas components shows all
used calibration gases and gives information
about the gas quality. With appropriate user
rights, you can also make changes here.
G a s Q u a l i t y A n a l y z e r
7
Page 59 of 135