8.3.5 Reset all parameters to default condition

If the device does not achieve acceptable results, e.g. by several incorrect settings, it may be
necessary to restore an older parameter setting, or the default condition
This procedure should only be used in emergency situations when all other
options have been exhausted. The current device data will be deleted. If in
doubt, contact Elster GmbH.
Be sure to save all previously required data on an independent external disk
or computer.
A backup according to chapter 8.1.3 must be available.
Working steps to put default setting on the device
1.
Start the program enSuite
2.
Back up all relevant data of the devices on an external disk or computer
Install the „Backup Parameterization" created according to chapter 8.1.3 on your PC
3.
4.
Connect with the device.
5.
Transfer the default parameterization in the device.
6.
The device will reboot and is now working with the factory settings.

8.4 EnSuite during sevice operation

8.4.1 Customize displays

In normal cases, the displays need not be edited or customized. But if you have the
authorization you can do the following (navigate to the shown enSuite page):
Page 108 of 135
Operation and parameter setting on the PC with enSuite
Figure 8.14: Basic display editor