Calibrate Menu - Xylem EXO Series User Manual

Advanced water quality monitoring platform
Table of Contents

Advertisement

KOR Software
4.3

Calibrate Menu

This menu is the main interface for calibrating EXO sensors. Calibration and verification settings will vary
by sensor. A device must be connected to access the Calibrate menu.
For detailed calibration procedures for individual sensors, reference Section 5 Calibration.
|
54
pg
Calibration menu overview
In the Calibrate menu, the list of installed sensors is shown
on the left side of the window. If your sonde is configured
with a depth sensor, Port D depth also appears in the list.
Sensor calibration menu
Select a sensor from the list to bring up a sensor-specific
menu of parameters. For example, selecting ODO (optical
dissolved oxygen) from the list, brings up a menu for
the enabled parameters ODO % sat and ODO mg/L. (To
change the enabled parameters, go to the Options | Units
menu.)
Click on a parameter to bring up the sensor-specific
calibration menu. The manufacturer recommended default
calibration parameters are in bold.
A typical calibration window shows 1, 2, or 3 calibration
points, depending on the sensor. If the sensor supports
only a single calibration point, then the other calibration
points are not active (grayed out).
Temperature: This field displays current temperature, if a
conductivity/temperature sensor is installed. If no sensor is
installed, user can input data into this field.
Barometer: Displays for DO calibrations only.
Standard Value: User-input field for the calibration
setpoint based on the value of the standard being used.