Xylem EXO Series User Manual page 16

Advanced water quality monitoring platform
Table of Contents

Advertisement

Location of Depth Sensor
(continued)
To assist with consistent horizontal orientation, the EXO2 sonde has
an indentation at the top of the sonde for a marker or positioning
pin.
The sonde should be installed with at least 1 cm of water above the
intake ports.
If a conductivity sensor is installed, the depth will be compensated
automatically for changes in the density of water as temperature
and salinity change.
Depth Configuration
EXO sondes must be ordered with a specific depth option: 0-10 m,
0-100 m, 0-250 m, or no depth. Once the depth selection is made,
the sonde's depth sensor cannot be changed.
|
16