Calibration; Why Calibrate; When Do You Have To Calibrate; Automatic Calibration (Autocal) - Xylem YSI pH 1310P Operation Manual

Ph meter
Table of Contents

Advertisement

TruLab pH 1310
ba76156e01
03/2014
5.2

Calibration

5.2.1 Why calibrate?

pH electrodes age. This changes the zero point (asymmetry) and slope of the
pH electrode. As a result, an inexact measured value is displayed. Calibration
determines and stores the current values of the zero point and slope of the
electrode.
Thus, you should calibrate at regular intervals.

5.2.2 When do you have to calibrate?

 After connecting a sensor
 Routinely within the framework of the company quality assurance
 When the calibration interval has expired

5.2.3 Automatic calibration (AutoCal)

Make sure that in the sensor menu, Buffer menu, the buffer set is correctly
selected (see 7.1.1 C
Use any one to five buffer solutions of the selected buffer set in ascending or
descending order.
Below, calibration with YSI buffers (YSI) is described. When other buffer sets
are used, other nominal buffer values are displayed. Apart from that, the proce-
dure is identical.
If single-point calibration was set in the menu, the calibration proce-
dure is automatically finished with the measurement of buffer solu-
tion 1 and the calibration record is displayed.
1.
Connect the pH electrode to the meter.
2.
Keep the buffer solutions ready.
When measuring without temperature sensor:
Temper the buffer solutions or measure the current temperature.
3.
In the measured value display, select the measured parameter pH or mV
with <M>.
4.
Start the calibration with <CAL>.
The calibration display for the first buffer appears (voltage display).
HANGING THE SETTINGS FOR P
pH value
H
,
MEASUREMENTS
PAGE
38).
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ysi ph 1310

Table of Contents