Following the results of the radiated immunity tests, if the device suffers radiated
electromagnetic disturbance, the normal functioning can be interrupted and a non-
authorized accessory change alarm will be displayed. In this case, the user only needs to
deactivate the alarm by pressing the OK button, make sure the electromagnetic
disturbance has disappeared and resume the treatment by selecting the corresponding
treatment parameters.
portable and mobile RF communication and the equipment and the equipments of the
CAPENERGY product range: C400, C300, C200, C100, C50 and C500.
Rated maximum
output power
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
Table J-4
Recommend separation distances between portable and mobile RF communication equipment
and the equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and C500.
Recommend separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
P
d=1,2
0,12
0,38
1,2
3,8
12
User Manual
Page 86 of 95
m
80 MHz to 800 MHz
P
d=1,2
0,12
0,38
1,2
3,8
12
800 MHz to 2,5 GHz
P
d=2,3
0,23
0,73
2,3
7,3
23