5.3 MAINTENANCE ROUTINES INTERVALS
In the following table you will find an overview of the necessary periodical maintenance
routines for C 400 devices (and the entire product range).
Frequency
Each 5 years.
After each
treatment
Before each
treatment
Monthly
Task
General revision of the equipment.
Verification
and
parameters of power, frequency and
temperature.
Clean the treatment accessories.
Inspection of the active plate's gel sleeve.
Inspection of the insulation of the active
plate and the integrity of the passive plate.
Inspection
of
the
capacitive electrodes.
Inspection of the accessories' cables and
connections.
Check the correct functionality of the
safety control
Inspection of the central unit.
User Manual
Page 41 of 95
checking
of
the
insulation
of
the
Executed by
MANUFACTURER
USER