Capenergy C 50 User Manual page 85

Diathermy equipment
Table of Contents

Advertisement

Guidance and manufacturer's declaration – electromagnetic immunity
The equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and C500 are intended for
use in the electromagnetic environment specified below. The customer or the user of the equipments of the
CAPENERGY product range: C400, C300, C200, C100, C50 and C500 should assure that it is used in such an
enviroment.
Immunity test
Conducted RF
3 Vrms
IEC 61000-4-6
of 150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
of 80 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the highest frequency range applies
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones
and land mobile radios, amateur radio, AMF and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and
C500 are used exceeds the applicable RF compliance level above, the equipments of the CAPENERGY
product range: C400, C300, C200, C100, C50 and C500 should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and C500.
b
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3V/m.
Table J-3. Electromagnetic immunity. Radiated emissions.
IEC 60601 test
level
User Manual
Compliance
Electromagnetic environment -
level
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part
CAPENERGY product
C300, C200, C100, C50 and C500,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter
Recommended separation distance
d=1.2
3 V
3 V/m
d= 1.2
The laboratory
limit is 1GHz
d=2.3
where P is the maximum output power
rating of the transmitter in watts (W)
according
manufacturer and d is the recommended
separation distance in meters (m).
Field
transmitters,
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Page 85 of 95
of the equipments of the
range:
P
P
80 MHz to 800 MHz
P
800 MHz to 2,5 GHz
to
the
transmitter
strengths
from
fixed
as
determined
C400,
RF
by
an