Guidance and manufacturer's declaration – electromagnetic immunity
The equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and C500 are
intended for use in the electromagnetic environment specified below. The customer or the user of the
equipments of the CAPENERGY product range: C400, C300, C200, C100, C50 and C500 should
assure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input-lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
Magnetic field
IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table J-2. Electromagnetic immunity
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
User Manual
Page 84 of 95
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for power
inputs/outputs
±1 kV line to line
±2 kV line to earth
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
70% U
T
(30% dip in U
)
T
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 s
3 A/m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the equipments of the
CAPENERGY product
range: C400, C300, C200,
C100, C50 and C500
requires continued
operation during power
mains interruptions, it is
recommended that the
equipments of the
CAPENERGY product
range: C400, C300, C200,
C100, C50 and C500 be
powered from an
uninterruptible power supply
or battery
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.