Capenergy C 50 User Manual page 58

Diathermy equipment
Table of Contents

Advertisement

VII. Do you wish to work with two(2) automatic electrodes (plates) and one (1)
insulated manual electrode (capacitive electrode holder)
VIII. Do you wish to work with three (3) automatic electrodes (plates) and one (1)
insulated manual electrode (capacitive electrode holder)
IX. Do you wish to work with one (1) non-insulated manual electrode (resistive
electrode holder)
X. Do you wish to work with one (1) automatic electrode (plate) and one (1) non-
insulated manual electrode (resistive electrode holder)
XI. Do you wish to work with two (2) automatic electrodes (plates) and one (1) non-
insulated manual electrode (resistive electrode holder))
XII. Do you wish to work with three (3) automatic electrodes (plates) and one (1) non-
insulated manual electrode (resistive electrode holder)
XIII. Do you wish to work with one (1) insulated manual electrode (capacitive electrode
holder) and one(1) non-insulated manual electrode (resistive electrode holder)
Before pressing the start button located at the back of the device:
1) When using any type of manual electrode holder: place the electrode on
the skin surface of the area you are going to treat.
2) When using automatic active plates: place the active plates on the
treatment area and make sure the contact between the plate and the skin is
ideal by means of elastic bandages, If you are going to use the automatic
plates without an operator being present, connect the safety control holder
which allows the patient to stop the treatment at any time.
Place the passive plate(s) on the area(s) that will receive the return current.
Program the treatment time and press the START button.
Adjust the power until the client perceives the generated heat level as pleasant.
At the end of the treatment, an acoustic signal will be heard. Press the STOP button
and set the power regulator to 0.
Remove the plates.
Clean the plates and/or the manual electrodes used during the treatment.
User Manual
Page 58 of 95

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

C 100C 200C 300C 400C 500

Table of Contents