Getting Started
WARNING. To avoid shock or burn hazard, use only the standard
accessories or accessories that are rated for the application. Use the
P5210 accessories with this product only.
Keep your fingers behind the finger guard on the probe body
whenever possible to reduce the risk of a shock from the circuit under
test.
Differential Inputs. The inputs are CAT II rated to a
maximum of 2,200 V
earth ground. They are also rated for a maximum
difference of 4,400 V
than 5,600 V
between a single input and ground.
Attenuation Range.
button sets the attenuation to 1,000X. In the lowered
position the range button sets the attenuation to 100X.
Use the 1,000X position for measurements up to a
maximum of 4,400 V
better signal resolution on connections below
440 V
Overrange Indicator. The overrange indicator lights red
if the voltage of the input signal exceeds the linear
operating range of the probe. When this happens, the
signal on the probe output does not accurately represent
the signal on the probe input.
Overrange Beeper.
button sets the overrange beeper to sound whenever the
overrange indicator lights.
Bandwidth Select.
button sets the full bandwidth (50 MHz). In the
lowered position the bandwidth is restricted to
approximately 5 MHz.
6
between either input and
RMS
between the inputs with less
RMS
) between the inputs or
(DC + peak AC)
In the raised position the range
. Use the 100X position for
RMS
.
RMS
In the raised position the overrange
In the raised position the bandwidth
P5210 Instructions