Tektronix P5210 Instructions Manual page 15

High-voltage differential probe
Hide thumbs Also See for P5210:
Table of Contents

Advertisement

Available languages

Available languages

CAUTION. To avoid damaging the input circuitry of the P5210 probe,
do not apply voltage that is more than 2,200 V
input and earth ground or more than 4,400 V
In addition, the peak voltage must be less than 5,600 V
between the inputs.
Above 1.5 MHz, the voltage limit decreases as frequency increases.
See Figure 2. The input limit applies to both the 100X and 1,000X
settings.
Category II maximum
voltage limit (2,200 V)
4000V
1000V
500V
Voltage
(RMS)
100V
50V
DC or 0
100k
Figure 2: Safety Limits (voltage between either input and earth ground
P5210 Instructions
RMS
RMS
RF burn risk area
(shaded)
500k
1M
5M
10M
Frequency (Hz)
Operating Basics
between either
between the inputs.
(DC + peak AC)
Voltage derating
with frequency
50M
)
13