Functional Check - Tektronix P5210 Instructions Manual

High-voltage differential probe
Hide thumbs Also See for P5210:
Table of Contents

Advertisement

Available languages

Available languages

Functional Check

To make a simple functional check of the P5210 probe, select a
source that supplies AC line voltage and use the following
procedure. This procedure verifies a majority of the circuitry within
the probe. (For a complete performance verification, refer to the
service manual that accompanies this product.)
1. Use the installation procedure starting on page 10 to connect the
output of the P5210 probe to a measurement instrument.
2. See Table 1 on page 11. Connect the inputs, set the range, and
perform the check as each line of Table 1 indicates.
Table 1: Functional Check
Range
Mode
Setting
Differential
100X or
1,000X
Common
100X or
Mode
1,000X
This completes the functional check procedure.
P5210 Instructions
Input 1
Input 2
(+ or –)
(– or +)
Check
Hot
Ground or
Measurement instrument
Neutral
displays or indicates the line
voltage
Hot (same
No signal
connection)
Getting Started
11