Cleaning, Disinfection and Sterilization
7.2

Cleaning Procedure

Cleaning is intended to remove all the visible soiling (e.g. organic and inorganic) from the
surface of the device and accessories so that the subsequent disinfection process can be
efficient.
®
Propex IQ
Apex Locator (and clamp, if used), Propex IQ
®
Smart IQ
Measurement Cable
Switch off Propex IQ
A
Propex IQ
Pay attention that no liquid enters the device connecting port, e.g. by keeping the cable
plugged-in.
Clean the devices by using a single-use wipe or soft cloth soaked in cleaning fluid
B
containing Alcohol Quaternary Ammonium active ingredients (e.g. CaviWipes™ or
CaviCide™).
For the Propex IQ
C
clean brush with soft bristle (e.g. nylon) slightly soaked in cleaning solution containing
Alcohol Quaternary Ammonium active ingredients (e.g. CaviWipes™ or CaviCides™).
Continue steps
D
and clean the brush as necessary.
®
Propex IQ
File Clip and Lip Clip
Pre-treatment of the file clip and lip clip – for personal protection
After use, place the file clip and lip clip directly into a dish filled with a suitable cleaning
and disinfection solution (Cidezyme, Johnson & Johnson Medical, 0.8% for at least
E
1 minute, maximum 2 hours) for cleaning, pre-disinfection and interim storage. Make
sure that the devices are fully immersed.
44/74
®
Apex Locator, then remove and discard the single-use
®
Disposable Barrier Sleeve (sleeves if the Propex IQ
®
Apex Locator, clean the grooves and device screws by using a
B
and
C
until no debris appear on the wipe / brush. Replace the wipe
Cleaning steps
®
Measurement Cable and X-
BENPPIQDFUWEB / Rev.04 / 08-2018
®