Dentsply Sirona Propex IQ Directions For Use Manual page 68

Apex locator
Hide thumbs Also See for Propex IQ:
Table of Contents

Advertisement

Appendix A: Electromagnetic Emissions and Immunity
Recommended separation distances between portable and mobile RF communications equipment
®
The Propex IQ
Apex Locator is intended for use in the electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the Propex IQ
Locator can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and the Propex IQ
Apex Locator as recommended below, according to the maximum output power of the
communication equipment.
Rated
maximum
output power of
150 kHz to 80 MHz
transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 3: An additional factor of 10/3 is used in calculating the recommended separation distance to
decrease the likelihood that mobile / portable communications equipment could cause interference
if it is inadvertently brought into patient areas.
68/74
and the Propex IQ
Separation distance according to frequency of transmitter
80 MHz to 800 MHz
×
=
1.2
d
P
0.12
0.38
1.2
3.8
12.0
®
Apex Locator
(m)
×
=
1.2
d
P
0.12
0.38
1.2
3.8
12.0
BENPPIQDFUWEB / Rev.04 / 08-2018
®
Apex
800 MHz to 2.5 GHz
×
=
2.3
d
P
0.23
0.73
2.3
7.3
23.0
®

Advertisement

Table of Contents
loading

Table of Contents