Cleaning, Disinfection and Sterilization

7
CLEANING, DISINFECTION AND STERILIZATION
7.1

General

You must consult the special Propex IQ
Guide before proceeding with cleaning disinfection and sterilization.
Disposable (single-use) Propex IQ
dental equipment / accessories or endodontic instruments. Always reprocess Propex IQ
Apex Locator.
The following parts shall be reprocessed before and after each treatment.
The reprocessing procedure might damage the (USB) plugs connected to the Propex IQ
Apex Locator / X-Smart IQ
during normal (plugged) use. Avoid reprocessing these plugs.
®
Propex IQ
Apex Locator
®
Propex IQ
Measurement Cable
®
X-Smart IQ
Measurement Cable
®
Propex IQ
File Clip
®
Propex IQ
Lip Clip
Dentsply Sirona accepts no liability in the event that these instructions are disregarded
or unvalidated processes are used to reprocess the accessories.
Use only properly maintained equipment and approved materials according to national
laws and/or regulations.
Always wear protective clothes for you own safety (gloves, mask, glasses and a
waterproof gown) for the reprocessing process.
BENPPIQDFUWEB / Rev.04 / 08-2018
®
Handpiece ports. They are unlikely to be contaminated
Part
®
Cleaning, Disinfection and Sterilization
®
Barrier Sleeves do not replace reprocessing of
Reprocessing steps
Clean and disinfect.
Do not sterilize.
Clean, Disinfect and Sterilize.
®
®
43/74