Step by Step Instructions
Color indication on
®
Propex IQ
Apex
Locator
32/74
Endodontic file progression
Audio signals (from the device) serve as indication of the file tip
progression towards reference point. The interval between the beeps
becomes shorter when the endodontic file approaches the reference
point for working length determination.
Region coronal to the reference point for working length
determination:
The endodontic file can be moved further apically.
LED: Flashing green.
Sound: Fast pulsating tone.
Reference point for working length determination:
This is the relevant position to determine the working length.
LED: Solid green.
Sound: Solid tone.
In combined mode, the Apical Reverse function can be activated.
®
Refer to Endo IQ
App User Manual – 6 Step by Step Instructions.
The position of the reference point can be adjusted with the "Shaping
Target" function, which is only available when Propex IQ
Locator is connected to the Endo IQ
App User Manual for instructions on that function.
Region beyond reference point:
Further file progression beyond the reference point is visualized.
LED: Solid green.
Sound: Solid tone.
®
®
App. Refer to the Endo IQ