Dentsply Sirona Propex IQ Directions For Use Manual page 71

Apex locator
Hide thumbs Also See for Propex IQ:
Table of Contents

Advertisement

Appendix A: Electromagnetic Emissions and Immunity
Input a.c. power port
PHENOMENON
Electrical fast
transients / bursts
a) l) o)
a) b) j) o)
Surges
Line-to-line
a) b) j) k) o)
Surges
Line-to-ground
Conducted
disturbances induced
c) d) o)
by RF fields
f) p) r)
Voltage dips
Voltage interruptions
f) i) o) r)
BENPPIQDFUWEB / Rev.04 / 08-2018
BASIC EMC
STANDARD OR TEST
METHOD
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
IMMUNITY TEST LEVELS
PROFESSIONAL
HEALTHCARE FACILITY
ENVIRONMENT
± 2 kV
100 kHz repetition
frequency
± 0,5 kV, ±1 kV
± 0,5 kV, ± 1 kV, ± 2 kV
m)
3 V
0,15 MHz - 80 MHz
m)
6 V
in ISM bands
between 0,15 MHz and
n)
80 MHz
e)
80% AM at 1 kHz
g)
0% UT; 0,5 cycle
At 0°, 45°, 90°, 135°, 180°,
q)
225°, 270° and 315°
0% UT; 1 cycle and
70% UT; 25/30 cycles
Single phase: at 0°
TEST RESULTS
Pass
Pass
Pass
Pass
Pass
h)
Pass
Pass
71/74