Dentsply Sirona Propex IQ Directions For Use Manual page 54

Apex locator
Hide thumbs Also See for Propex IQ:
Table of Contents

Advertisement

Troubleshooting
®
Propex IQ
Apex Locator does not switch on or it shuts off during treatment.
Possible causes
The battery is completely empty (all LEDs
are OFF).
The charger is not connected properly into
the power socket.
The battery is wrongly inserted.
The original battery has been replaced
with a non-original one.
ON/OFF/Standby button malfunction.
Electronic malfunction.
54/74
Solutions
Connect the device to the charger and
charge the device.
Ensure that the charger is plugged in
correctly and the power socket is under
tension/switched ON (if applicable).
Ensure that the original adapter is being
used.
Open the battery compartment and check
the battery orientation.
Try pressing the ON/OFF/Standby button
several times.
Replace the batteries with a new original
battery.
Perform a factory reset.
Contact Dentsply Sirona Endodontics
Customer Service.
Send the device to Dentsply Sirona
Endodontics Customer Service.
BENPPIQDFUWEB / Rev.04 / 08-2018