Apex locator
Hide thumbs Also See for Propex IQ:
Table of Contents

Advertisement

Step by Step Instructions
After reprocessing of Propex IQ
Cleaning, Disinfection and
use Propex IQ
top. If you are using the optional Propex IQ
second sleeve over the clamp as follows:
A
B
Press the ON/OFF/Standby button on Propex IQ
C
Connect the Propex IQ
D
Connect the Propex IQ
E
Connect the Propex IQ
F
Perform a closed-loop test. See
G
Put the lip clip (3) on the patient's lip on the opposite site of the tooth to be treated.
Insert the endodontic file (5) into the root canal and attach the file clip (4) to the metallic
H
part of the endodontic file (5) directly below the plastic handle and above the stopper
(6).
Move the endodontic file apically until reference point (default reference point or
I
reference point defined as per Shaping Target function) is reached.
J
Set the silicon stopper at the coronal reference.
Remove the endodontic file from the root canal and use an external ruler to measure the
K
distance between the tip of the endodontic file and the silicon stopper. This information
will be relevant for subsequent files in your sequence.
Compare this distance information with the corresponding information given by pre-
L
treatment radiographs.
Proceed with endodontic treatment. Carefully monitor file progression to avoid over-
M
instrumentation.
34/74
®
Apex Locator and its accessories (see Section
Sterilization), place Propex IQ
®
Disposable Barrier Sleeve, with the long end of the barrier sleeve at the
®
Measurement Cable (2) to the device (1).
®
Lip Clip (3) to the Propex IQ
®
File Clip (4) to the Propex IQ
6.9.1 Closed-loop
Action
®
®
Clamp on the back of the device, pull a
®
Apex Locator (1).
®
Measurement cable (2).
®
Measurement Cable (2).
test.