Cleaning, Disinfection and Sterilization

7
Cleaning, Disinfection and Sterilization
7.1

Foreword

Regular maintenance of X-SMART IQ™ is essential to ensure that work
performed is completely hygienic and that the device will have a long
service life.
The components of X-SMART IQ™ have not been disinfected or
sterilized when they are delivered. Please perform the necessary steps
before the first treatment.
You are responsible for the sterility of accessories; always ensure that
only validated methods of cleaning, disinfection and sterilization are
used, that regular maintenance and inspections are performed on the
equipment used (disinfector, sterilizer) and that the validated
parameters are maintained in every cycle.
Comply with your national guidelines, standards and requirements for
cleaning, disinfection and sterilization.
Do not place the X-SMART IQ™ motor handpiece in an autoclave
unit or washer-disinfector.
Wear protective clothing during all cleaning, disinfection and
sterilization processes.
7.2

Single Use of Custom Barrier

The use of Custom Barrier over the motor handpiece is mandatory.
The barrier is designed for single use and must be disposed of after
use in accordance to territorial applicable regulations.
Please refer to the instructions for use on the Custom Barrier box.
7.3

Disinfection of Components

The components listed on
wipe-down with a disinfection wipe prior to and following each
treatment. They are not approved for machine-based cleaning (i.e.
Thermal disinfection, washer disinfectors) or sterilization.
B EN XSIQ DFU MAS / Rev.10 / 01-2017
page 55