Operating Instructions Safety Cabinet Safe 2020 / Maxisafe 2020
5.
Start-up
5.5

RS 232 interface connection

The RS 232 interface has been designed for a cable connection
with 9-pin connectors and a contact assignment of 1:1.
Connection of the device:
1. Turn PC off.
2. Fig. 11: Connect the connector of the serial interface
cable (not comprised in the scope of delivery) to the
socket [1] at the supply interface at the front of the
electrical box.
3. Connect the serial interface cable to an unassigned
slot COM 1/COM 2 etc. at the PC.
4. Turn PC on.
Transfer protocol:
The interface must be configured as follows:
Baud:
Data bits:
Parity:
Stop bit:
Protocol:
FIFO–puffer(extended modulation):
Occupancy of conductors:
Type of connector [X] : 9-Pin SUB-D
Pin 2: TxD
Pin 3: RxD
Pin 5: GND
5.6

UV connection

Fig. 12: UV disinfection adapter (optional) for an external UV
disinfection unit.
Voltage:
Current:
Connectors:
30
9600
8
none
1
none
enabled
230 V
max. 1,1 A
[1], [2], [3] and PE-sign
Fig. 11
RS 232 interface connection
Fig. 12
UV connection